Force 40 Electrosurgical Unit
- Valley Lab Force 40 ESUSpecifications:•Automatic activation impedance 40-100ohms•Automatic deactivation impedance: <20ohms >2000ohms•Time delay range: .1 to 5.1 seconds•Activation tone: 45 to 65 dBA•Alarm Tone: >65 dBA•Weight: 20lb (9.1 kg)•Dimensions: 7.8 in H x 16.1 in W x 11.9 in D(199mm x 409 mm x 303 mm)•Operates on 110/240V 50/60Hz.Output Waveform:•Cut: 500kHz sinusiod•Blend 1: 500kHz bursts of sinusiod @50% duty cycle recurring @ 31.25 kHz•Blend 2: 500kHz bursts of sinusiod @37.5% duty cycle recurring @ 31.25 kHz•Desiccate: 500kHz damped sinusiodal burst with a repetition frequency of 250 kHz•Fulgurate: 500kHz damped sinusiodal burst with a repetition frequency of 62.5 kHz•Spray: 500kHz damped sinusiodal burst with a repetition frequency of 31.25 kHz•Standard Bipolar: 500kHz sinusiod•Precise Bipolar: 500kHz sinusiodOutput Characteristics•Cut: Max P-P 3300 Rate Load 300 Max Power 300 Crest Factor 2.1 @ 100W•Blend 1: Max P-P 3800 Rate Load 300 Max Power 250 Crest Factor 2.1 @ 100W•Blend 2: Max P-P 4000 Rate Load 300 Max Power 200 Crest Factor 2.1 @ 100W•Desiccate: Max P-P 3500 Rate Load 300 Max Power 200 Crest Factor 2.1 @ 100W•Fulgurate: Max P-P 6500 Rate Load 300 Max Power 150 Crest Factor 2.1 @ 100W•Spray: Max P-P 9000 Rate Load 300 Max Power 150 Crest Factor 2.1 @ 100W•Standard Bipolar: Max P-P 550 Rate Load 100 Max Power 50 Crest Factor 2.1 @ 100W•Precise Bipolar: Max P-P 550 Rate Load 100 Max Power 50 Crest Factor 2.1 @ 100W
VALLEYLAB Force 40
- ID: 1121
- Offer ID# 1121
- Product Code: VALLEYLAB Force 40
- Availability: In Stock